Separation of Explosives from Drinking Water


Column Purospher® RP-18 HC LiChroCART® 250-4, 5 μm
Mobile phase A: Acetonitrile / Methanol; 20 / 80, v/v
B: Sodium dihydrogenphosphate buffer (c = 0.01 mol/L, pH 4.5)
Gradient 0 min 35 % A; 28 min 55 % A; 40 min 85 % A; 50 min 85 % A; 51 min 35 % A; 71 min 35 % A
Flow rate 0.8 mL/min
Detection DAD 210 and 235 nm
Temperature 36 °C
Inj. volume 40 μL
Sample 1. Octogen (HMX)
2. Picric acid
3. Hexogen (RDX)
4. Ethylene Glycol Dinitrate (EGDN)
5. Diethylene Glycol Dinitrate (DEGN)
6. 1,3,5-Trinitrobenzene
7. 1,3-Dinitrobenzene
8. Tetryl
9. Nitroglycerine

10. 2,4,6-Trinitrotoluene
11. 4-Amino-2,6-dinitrotoluene
12. 2-Amino-4,6-dinitrotoluene
13. 2,6-Dinitrotoluene
14. 2,4-Dinitrotoluene
15. 2-Nitrotoluene
16. 4-Nitrotoluene
17. 3-Nitrotoluene
18. Nitropenta (PETN)
19. Hexyl
Merck:/Freestyle/LE-Lab-Essentials/Chromatography/Diagrams/LE-Chrom-Purospher HPLC-Purospher RP-18 HC HPLC Columns-Separation of explosives from drinking water-480x478-20140214.png