  • Automatic On-line Solid-phase Extraction-Electrothermal Atomic Absorption Spectrometry Exploiting Sequential Injection Analysis for Trace Vanadium, Cadmium and Lead Determination in Human Urine Samples.

Automatic On-line Solid-phase Extraction-Electrothermal Atomic Absorption Spectrometry Exploiting Sequential Injection Analysis for Trace Vanadium, Cadmium and Lead Determination in Human Urine Samples.

Analytical sciences : the international journal of the Japan Society for Analytical Chemistry (2015-05-12)
Georgia Giakisikli, Alejandro Ayala Quezada, Junpei Tanaka, Aristidis N Anthemidis, Hiroya Murakami, Norio Teshima, Tadao Sakai
ABSTRACT

A fully automated sequential injection column preconcentration method for the on-line determination of trace vanadium, cadmium and lead in urine samples was successfully developed, utilizing electrothermal atomic absorption spectrometry (ETAAS). Polyamino-polycarboxylic acid chelating resin (Nobias chelate PA-1) packed into a handmade minicolumn was used as a sorbent material. Effective on-line retention of chelate complexes of analytes was achieved at pH 6.0, while the highest elution effectiveness was observed with 1.0 mol L(-1) HNO3 in the reverse phase. Several analytical parameters, like the sample acidity, concentration and volume of the eluent as well as the loading/elution flow rates, have been studied, regarding the efficiency of the method, providing appropriate conditions for the analysis of real samples. For a 4.5 mL sample volume, the sampling frequency was 27 h(-1). The detection limits were found to be 3.0, 0.06 and 2.0 ng L(-1) for V(V), Cd(II) and Pb(II), respectively, with the relative standard deviations ranging between 1.9 - 3.7%. The accuracy of the proposed method was evaluated by analyzing a certified reference material (Seronorm(TM) trace elements urine) and spiked urine samples.

MATERIALS
Product Number
Brand
Product Description

Sigma-Aldrich
Ammonium acetate, Molecular Biology, ≥98%
Sigma-Aldrich
Ammonium acetate, BioXtra, ≥98%
Sigma-Aldrich
Ammonium acetate solution, Molecular Biology, 7.5 M
Sigma-Aldrich
Ammonium acetate, reagent grade, ≥98%
Sigma-Aldrich
Nitric-14N acid solution, ~10 N in H2O, 99.99 atom % 14N
Sigma-Aldrich
Ammonium acetate, 99.999% trace metals basis
Sigma-Aldrich
Nitric acid, ACS reagent, ≥90.0%
Sigma-Aldrich
Nitric acid, 70%, purified by redistillation, ≥99.999% trace metals basis