- Screw the 14 L/min flow-limiting orifice into the filter holder outlet and connect the holder to a vacuum source (of at least 55 cm Hg at 14 L/min) using a hose.
- Carefully place a Type AA black filter, grid side up, on the support screen in the filter holder base using clean forceps. Similarly place a Type SM white filter on top of the filter. Lock in place with the blue locking ring. When ready to sample, see step 3.
|
Filter is centered on base of garment monitoring holder. The prefilter is placed onto the filter.
|
- Place the outer surface of the test garment over the tapered (male) body of the holder. Secure the prefilter adapter assembly over the test portion of the fabric.
- Apply vacuum and sample with the same filter for a period of one minute for each of the five test areas. Be sure to turn vacuum off after each test.
- Open the filter holder, remove the sample filter, and place it on a PetriSlide™ device for subsequent counting. (Refer to Analysis Techniques section)
|
Test garment secured between prefilter adapter section (right), and the filter holder (left).
|