Flow Cytometry Applications - Bioprocess |
Request Information |
A wide range of easyCyte™ systems is now able to address the demands of bioprocess. The easyCyte 5HT HPL is uniquely suited for simple viability determinations and precise cell counts, while the easyCyte™ 8HT provides additional capabilities for more complex analyses. To learn more about instruments and applications suited for bioprocess, please read the selections below.
Count & Viability Testing
A recent internal study validates this method for bioprocess development or process monitoring (see Technical Brief TB2134EN00).
Recovery versus Trypan Blue | Expected cell concentration | Total cell concentration recovery (%) | Viability Recovery (%) |
---|---|---|---|
CHO | 10e5 | 99.5 | 96.1 |
10e7 | 101.9 | 101.8 | |
ANJ2 | 10e5 | 111.9 | 94.4 |
10e7 | 103.4 | 90.9 | |
VERO | 10e5 | 101.6 | 96.4 |
10e7 | 98.3 | 90.5 | |
Average | 102.8 | 95.0 | |
Standard Deviation | 4.8 | 4.2 |
Viability and cell concentrations were tested with Viacount® for three separate bioprocess cell lines on the guava easyCyte™ HT HPL. As shown above, results were well within the range of acceptance criteria and showed concordance with expected cell concentrations of all samples tested.
EasyCyte™ 5HT hpl | |||||
---|---|---|---|---|---|
CV (%) | Expected cell concentration | Total cell concentration CV (%) | CV: difference between high cell abd low cell concentration | Viability CV (%) | CV: difference between high cell and low cell concentration |
CHO | 10e5 | 4.4 | .09 | 1.4 | -1.2 |
10e7 | 3.5 | 2.6 | |||
ANJ2 | 10e5 | 6.6 | 3.5 | 1.6 | -4.8 |
10e7 | 3.1 | 6.4 | |||
VERO | 10e5 | 6.2 | 2.2 | 1.2 | -5.2 |
10e7 | 4 | 6.4 | |||
Trypan blue | |||||
CHO | 10e5 | 17.3 | 5.1 | 1.2 | -4.4 |
10e7 | 12.2 | 5.6 | |||
ANJ2 | 10e5 | 29.1 | 14.8 | 3.1 | -13.6 |
10e7 | 14.3 | 16.7 | |||
VERO | 10e5 | 19.8 | 9.2 | 1.1 | -11.7 |
10e7 | 10.6 | 12.8 |
CVs were evaluated for viability and cell concentration results across three different bioprocess cell lines at low and high concentrations. Guava easyCyte™ results show the best agreement with the acceptance criteria of the study as compared to Trypan Blue. Differences of CV between high and low concentration for Guava are lower than differences of CV of Trypan Blue for total cell concentration and viability.
Description of the easyCyte™ 5HT HPL system used in this bioprocess application are listed here:
The ViaCount® Assay analyzes cells based on differential permeabilities of two DNA-binding dyes. The nuclear dye stains only nucleated cells, while the viability dye brightly stains dying cells. Debris is excluded from results based on negative staining with the nuclear dye.