QuestionAnswer
What is the recommended rotor for Amicon Ultra units?Any rotor capable of accommodating 22.7 cm in diameter and 124 mm length centrifuge tubes for the Amicon Ultra-4 and 35 mm in diameter and 121 mm in length for the Amicon Ultra-15 can be used for Amicon Ultra units.
How should I select the appropriate Molecular Weight Cutoff (MWCO) for Amicon Ultra?Select a Molecular Weight Cutoff that is 2x smaller that the Molecular Weight (MW) of the solute of interest.
Which membranes are available in Amicon Ultra units?Amicon Ultra units are only available with Ultracel regenerated cellulose material. This high recovery/low binding membrane is offered in 3,000, 10,000, 30,000, 50,000 & 100,000 MWCO.
Can the Amicon Ultra be spun to dryness?No, the Amicon Ultra has a built in deadstop.
What is the chemical compatibility of Amicon Ultra?Amicon Ultra units are intended for use with biological fluids and aqueous solutions. Please refer to the user guide, catalogue, or contact Technical Support.
Do you have any centrifugal filter units that are pyrogen free?We do not have any centrifugals devices that we claim are pyrogen free. Even the sterile devices could contain pyrogens.