The small dilutions are required to improve recovery in serum/plasma samples due to a strong matrix effect. Improved recovery was observed when using 1:6 diluted samples compared to neat or lower dilution. It may seem counter-intuitive, but the dilution reduces the matrix effect and improves sample recovery and detectability.
HCCBP1MAG-58K
MILLIPLEX® Human Circulating Cancer Biomarker Magnetic Bead Panel - Cancer Multiplex Assay
Circulating Cancer Biomarker Bead-Based Multiplex Assays using the Luminex technology enables the simultaneous analysis of multiple oncology biomarkers in various tumor types in human serum, plasma and cell culture samples.
Sinonimo/i:
human cancer biomarker immunoassay panel, luminex human circulating cancer biomarkers multiplex assay, millipore human circulating cancer biomarkers multiplex kit
About This Item
Vai a
Livello qualitativo
Reattività contro le specie
human
Produttore/marchio commerciale
Milliplex®
assay range
accuracy: 35-108%
standard curve range: 0.09-67 mU/mL
(β-HCG)
standard curve range: 0.7-500 pg/mL
(IL-6)
standard curve range: 1.3-1000 pg/mL
(IL-8, TNFα)
standard curve range: 2.7-2000 pg/mL
(TRAIL, TGF-α)
standard curve range: 6.9-5000 pg/mL
(SCF)
tecniche
multiplexing: suitable
Metodo di rivelazione
fluorometric (Luminex xMAP)
Condizioni di spedizione
wet ice
1 of 4
Questo articolo | HCCBP3MAG-58K | HCCBP2MAG-58K | HAGP1MAG-12K |
---|---|---|---|
species reactivity human | species reactivity human | species reactivity human | species reactivity human |
assay range accuracy: 35-108%, standard curve range: 0.7-500 pg/mL | assay range accuracy: 74-126%, standard curve range: 0.004-3 ng/mL | assay range accuracy: 92-107%, standard curve range: 0.01-50 ng/mL | assay range accuracy: 42-131%, standard curve range: 1.4-1,000 pg/mL |
manufacturer/tradename Milliplex® | manufacturer/tradename Milliplex® | manufacturer/tradename Milliplex® | manufacturer/tradename Milliplex® |
Quality Level 200 | Quality Level 200 | Quality Level 200 | Quality Level 200 |
shipped in wet ice | shipped in wet ice | shipped in wet ice | shipped in wet ice |
detection method fluorometric (Luminex xMAP) | detection method fluorometric (Luminex xMAP) | detection method fluorometric (Luminex xMAP) | detection method fluorometric (Luminex xMAP) |
Descrizione generale
MILLIPLEX® Circulating Cancer Biomarker Panel 1 is a 25-plex kit that can be used for the simultaneous quantification of any or all of the following analytes in serum, plasma and tissue/cell lysate and culture supernatant samples: AFP, CA125, CA15-3, CA19-9, CEA, CYFRA21-1, sFas, sFasL, FGF2, β-HCG, HE4, HGF, IL-6, IL-8, Leptin, MIF, Osteopontin, Prolactin, PSA (free), PSA (total), SCF, TGFα, TNFα, TRAIL and VEGF. Note that certain biomarkers are tumor specific, such as PSA, while others, such as IL-8, have been detected in many cancers. PSA (free) and PSA (total) cannot be plexed together. This kit uses a 96-well format, contains a lyophilized standard cocktail, two internal assay quality controls and can measure up to 38 samples in duplicate.
The Luminex® xMAP® platform uses a magnetic bead immunoassay format for ideal speed and sensitivity to quantitate multiple analytes simultaneously, dramatically improving productivity while conserving valuable sample volume.
Panel Type: Circulating Cancer
Applicazioni
- Analytes: α-Fetoprotein (AFP), CA125, CA15-3, CA19-9, CEA, CYFRA21-1, sFas (TNFRSF6), sFasL, FGF-2 (FGF-basic), β-HCG (HCGβ), HE4, HGF, IL-6, IL-8 (CXCL8), Leptin, MIF, Osteopontin (OPN), Prolactin, PSA (free)*, PSA (total)*, SCF, TGFα, TNFα, TRAIL (TNFSF10), VEGF.
- Note: Total PSA and Free PSA cannot be plexed together
- Recommended Sample Type: Human serum, plasma or tissue/cell culture supernatants or lysates
- Recommended Sample Dilution: 25 μL per well of 1:6 diluted plasma or serum, or 25 μL cell culture psaesaa sample undiluted or diluted in appropriate control medium
- Assay Run Time: Overnight (16-18 hr) at 2-8°C
- Research Category: Cancer
Azioni biochim/fisiol
Cross-reactivity between the antibodies and any of the other analytes in this panel is non-detectable or negligible.
Caratteristiche e vantaggi
Confezionamento
Nota sulla preparazione
Altre note
Human Circulating Cancer Biomarker Panel 1 Quality Controls 1 and 2 (HCC-6058)
Serum Matrix (HCP-SM)
Note: Contains 0.08% Sodium Azide
Set of one 96-Well Plate with 2 Sealers (MAG-PLATE)
Assay Buffer (L-AB)
Note: Contains 0.08% Sodium Azide
10X Wash Buffer (L-WB)
Note: Contains 0.05% Proclin
Human Circulating Cancer Biomarker Panel 1 Detection Antibodies (HCC-1058 or HCC-1058-2)
Streptavidin-Phycoerythrin (L-SAPE 4)
Bead Diluent (LBD-4)
Note legali
Esclusione di responsabilità
Avvertenze
Danger
Indicazioni di pericolo
Classi di pericolo
Acute Tox. 4 Dermal - Acute Tox. 4 Inhalation - Acute Tox. 4 Oral - Aquatic Chronic 2 - Eye Dam. 1 - Skin Sens. 1 - STOT RE 2
Organi bersaglio
Respiratory Tract
Codice della classe di stoccaggio
10 - Combustible liquids
Certificati d'analisi (COA)
Cerca il Certificati d'analisi (COA) digitando il numero di lotto/batch corrispondente. I numeri di lotto o di batch sono stampati sull'etichetta dei prodotti dopo la parola ‘Lotto’ o ‘Batch’.
Possiedi già questo prodotto?
I documenti relativi ai prodotti acquistati recentemente sono disponibili nell’Archivio dei documenti.
-
Why is there a 1:6 sample dilution in serum matrix in the protocol for the Human Cancer Biomarker Panel 1 Milliplex kit (HCCBP1MAG-58K)?
1 risposta-
Utile?
-
Filtri attivi
Il team dei nostri ricercatori vanta grande esperienza in tutte le aree della ricerca quali Life Science, scienza dei materiali, sintesi chimica, cromatografia, discipline analitiche, ecc..
Contatta l'Assistenza Tecnica