Our broad portfolio consists of multiplex panels that allow you to choose, within the panel, analytes that best meet your needs. On a separate tab you can choose the premixed cytokine format or a single plex kit.
Cell Signaling Kits & MAPmates™
Choose fixed kits that allow you to explore entire pathways or processes. Or design your own kits by choosing single plex MAPmates™, following the provided guidelines.
The following MAPmates™ should not be plexed together:
-MAPmates™ that require a different assay buffer
-Phospho-specific and total MAPmate™ pairs, e.g. total GSK3β and GSK3β (Ser 9)
-PanTyr and site-specific MAPmates™, e.g. Phospho-EGF Receptor and phospho-STAT1 (Tyr701)
-More than 1 phospho-MAPmate™ for a single target (Akt, STAT3)
-GAPDH and β-Tubulin cannot be plexed with kits or MAPmates™ containing panTyr
.
Catalogue Number
Ordering Description
Qty/Pack
List
This item has been added to favorites.
Select A Species, Panel Type, Kit or Sample Type
To begin designing your MILLIPLEX® MAP kit select a species, a panel type or kit of interest.
Custom Premix Selecting "Custom Premix" option means that all of the beads you have chosen will be premixed in manufacturing before the kit is sent to you.
Catalogue Number
Ordering Description
Qty/Pack
List
This item has been added to favorites.
Species
Panel Type
Selected Kit
Qty
Catalogue Number
Ordering Description
Qty/Pack
List Price
96-Well Plate
Qty
Catalogue Number
Ordering Description
Qty/Pack
List Price
Add Additional Reagents (Buffer and Detection Kit is required for use with MAPmates)
Qty
Catalogue Number
Ordering Description
Qty/Pack
List Price
48-602MAG
Buffer Detection Kit for Magnetic Beads
1 Kit
Space Saver Option Customers purchasing multiple kits may choose to save storage space by eliminating the kit packaging and receiving their multiplex assay components in plastic bags for more compact storage.
This item has been added to favorites.
The Product Has Been Added To Your Cart
You can now customize another kit, choose a premixed kit, check out or close the ordering tool.
AFS® D Water Purification System
Cost-effective solution for analyzers with degassed pure water needs of up to 320 Liters dailyMore
Cost-effective solution for analyzers with degassed pure water needs of up to 320 Liters daily Less
C73374Cost-effective solution for analyzers with degassed pure water needs of up to 320 Liters daily
CLRW quality
AFS® D Systems provide a cost-effective, low maintenance water purification solution for analyzers with degassed pure water needs up to 320 L daily. AFS® D systems combine complementary water purification techniques; Complete pretreatment, advanced reverse osmosis (RO), Q-Gard® polishing pack, exposure to a bactericidal UV lamp, degassing, and final 0.22 µm filtration ensure a reliable and consistent water quality. Water produced by the systems complies with the Clinical and Laboratory Standards Institute® (CLSI) guideline for clinical laboratory reagent water (CLRW). Depending on daily volume needs, different production flow rates are available (8 and 16 L/h).
The systems are designed to fulfill accreditation needs (such as CAP™ 15189 accreditation to the ISO® 15189:2007 standard), and have full monitoring and water quality archiving capabilities that allow storage of up to six months of data. Activation of Millitrack® Essential e-Solution allows easy, rapid access to this information and also enhances data management control and provides remote system access.
User friendly & easy maintenance
Robust AFS® D systems require little maintenance, which means less analyzer downtime. Systems have a small footprint and can be installed wherever it is convenient in the laboratory: on a wall, on or under the bench, or on a cart. AFS® D systems are customizable with a wide range of options and accessories, including reservoirs, a sanitization kit and an aeration kit. Optimized lifetimes for the systems’ pretreatment and polishing packs, as well as low feed water consumption make AFS® D systems a cost-effective choice.
Water Specifications
AFS® D Essential CLRW water (CLSI Clinical Laboratory Reagent Water)
Resistivity at 25 °C
> 15 MΩ•cm
TOC
Typically < 30 ppb
Bacteria
< 1 CFU/mL
Silica
< 5 µg/L
Dissolved oxygen
Typically 6 - 8 ppm
System
Permeate Flow Rate
Peak Dispense Rate (to atmosphere)
Dispensing pressure to analyzer
AFS® 8D
8 L/h
Up to 0.6 L/min*
1 bar ± 10 % (13-16 psi)*
AFS® 16D
16 L/h
Up to 0.6 L/min*
1 bar ± 10 % (13-16 psi)*
* Dispensing to atmosphere (without any flow restriction due to the analyzer)
Feed water
Feed water quality
Potable
Feed water pressure
1 - 6 bar
Feed water temperature
5 - 35 °C
Minimum flow rate feed
> 5 L min at 2 bar
Total chlorine *
< 3 ppm
Langelier Saturation Index
< 0.3
Maximum Fouling Index (SDI)
< 12
Tap water connection
½” Gaz M
* For feed water with total chlorine concentration between 1 ppm and 3 ppm, Merck Millipore suggests the use of an external PrePak.
System Specifications
Dimensions (H x W x D)
470 x 268 x 426 mm (18.5 x 10.6 x 16.8 in)
Operating weight
< 22.2 kg (49 lb)
Electric power supply voltage
100 - 240 V +/- 10%
Electric power supply frequency
50 - 60 Hz +/- 10%
Power consumption
200 W or 250 VA
AFS® D main unit data connection
Ethernet (RJ45)
Safety
The AFS® D system is tested by an independent and accredited company for compliance with the CE directives related to safety and electromagnetic compatibility. The report can be consulted on demand. The AFS® D system is built using components and practices recommended by UL and has been cUL marked. The registration can be verified on the UL web site ( http://www.ul.com).