Analysis of small acylcarnitines using Ascentis® Express ZIC®-HILIC HPLC column

Sakur Mahmud1, Arun Babu Kumar1

Merck 811 Paloma drive, Round Rock, TX, USA

Abstract

Acylcarnitine profiling is an important analytical approach for the detection of metabolic abnormalities and is used in newborn screening. This application note demonstrates LC-MS/MS analysis of 13 stable isotope-labeled acylcarnitines (A-206) using an Ascentis® Express ZIC®-HILIC HPLC column (50 × 3.0 mm, 2.7 μm). The 13 target analytes were detected within a 10-minute chromatographic run using dMRM detection, with chromatographic separation achieved across the evaluated concentration range. Calibration curves showed strong concentration-response relationships, while the evaluated method performance included determination of LOD, LOQ, precision, and peak shape for all target analytes. The results demonstrate the application of the Ascentis® Express ZIC®-HILIC column for LC-MS/MS analysis of the evaluated stable isotope-labeled acylcarnitine mix. 

Section overview

Experimental

Materials and methods

A 13-component stable isotope-labelled acylcarnitine 13 component mix (A-206) was used for analysis. All other reagents were LC-MS grade. The HPLC conditions used for the analysis are summarized in Table 1, and the corresponding MS conditions for dMRM analysis are provided in Table 2. Tables 3 and 4 summarizes the dilution factors, nominal levels, and corresponding concentrations of the isotope-labeled acylcarnitines at each nominal level, with methanol used as the diluent.

The following column and instrument parameters were used for the experiments:

  • Ascentis® Express ZIC®-HILIC column 50 x 3 mm, 2.7 µm
  • Agilent 1290 HPLC system coupled with Agilent Ultivo Triple Quadrupole (QQQ) Mass Spectrometer

Results and discussion

The chromatographic separation of the 13 isotope-labeled acylcarnitines was evaluated using the Ascentis® Express ZIC®-HILIC HPLC column and dMRM detection. Figure 1 shows the dMRM chromatogram obtained for the (A-206) acylcarnitine mix at the evaluated concentration level. All 13 target analytes were detected within the chromatographic run. Although the earliest-eluting analytes showed partial chromatographic overlap, individual dMRM transitions were monitored for each analyte, providing MS/MS selectivity for their analysis. Resolution values ranged from 0.9 to 5.6, while tailing factors ranged from 0.9 to 1.5 at the nominal concentration (standard G).

LODs and LOQs were determined for each analyte from linear regression plots of an eight-point calibration curve generated across the evaluated concentration range. Figure 2. illustrates the regression plots for the most and least abundant analytes. The calibration curves showed strong concentration-response correlation across the dilution range, with R² values exceeding 0.99 for all 13 analytes. The LODs ranged from 3.4 to 106.2 ng/mL, and the LOQs ranged from 11.4 to 354.0 ng/mL. RSD values of absolute area at nominal concentration ranged from 1.1 to 3.8% (Table 4).

dMRM chromatogram of a 13-component stable isotope-labeled acylcarnitine standard mix (Standard G) analyzed using an Ascentis® Express ZIC®-HILIC HPLC column. All 13 analytes were detected within a 10-minute chromatographic run, with varying degrees of separation between adjacent peaks. Individual dMRM transitions enabled selective monitoring of each target analyte.

Figure 1.dMRM chromatogram of the 13-component stable isotope-labeled acylcarnitine mix obtained using an Ascentis® Express ZIC-HILIC HPLC column at nominal concentration (Standard G). All 13 target analytes were detected within the 10-minute chromatographic run, with varying degrees of chromatographic separation between adjacent peaks. Individual dMRM transitions enabled selective monitoring of each target analyte.

Linear regression calibration plot for iC5:0 d9 showing the relationship between concentration (ng/mL) and abundance (×1000). The calibration curve exhibits linearity with the regression equation y = 715.54x − 420.78 and a coefficient of determination (R²) of 0.9993.
Linear regression calibration plot for C2 d3 showing the relationship between concentration (ng/mL) and abundance (×1000). The calibration curve exhibits linearity with the regression equation y = 559.64x − 6048.8 and a coefficient of determination (R²) of 0.9981.

Figure 2. Linear regression plots for iC5:0 d9 and C2 d3, representing analytes with lower and higher concentration ranges, respectively, in the 13-component mix.

The Table 5 summarizes the analytical performance characteristics of the 13 isotopically labeled acylcarnitine standards, including limits of detection (LODs; 3σ/slope), limits of quantification (LOQs; 10σ/slope), precision (n = 5 at the nominal concentration), and chromatographic suitability of the developed LC-MS/MS method for quantitative analysis. Here, σ represents the standard deviation of residuals, and the slope was obtained from the linear regression equation for each analyte across all eight calibration levels. Residuals at each calibration level were calculated as the difference between the fitted and experimental responses. The residual standard deviation was calculated as the square root of the sum of squared residuals (SSE) divided by the degrees of freedom (DF): SSE n − 2 , where SSE is the sum of squared residuals and DF is the degrees of freedom, calculated as n-2, with n representing the number of calibration standards.

Conclusion

The Ascentis® Express ZIC®-HILIC (50 × 3 mm, 2.7 µm) column was evaluated for the LC-MS/MS analysis of 13 stable isotope-labeled acylcarnitines using dMRM detection. All 13 target analytes were detected within the 10-minute chromatographic run, with resolution values ranging from 0.9 to 5.6 and tailing factors from 0.9 to 1.5 at nominal concentration. LODs ranged from 3.4 to 106.2 ng/mL, LOQs from 11.4 to 354.0 ng/mL, and RSD values at the nominal concentration from 1.1 to 3.8%. The results demonstrate the application of the Ascentis® Express ZIC®-HILIC HPLC column (50 × 3.0 mm, 2.7 μm) for chromatographic separation and LC-MS/MS analysis of the evaluated acylcarnitine mix.

Related products

Reagents, solvents and standards

Loading

References

1.
Houten SM, Violante S, Ventura FV, Wanders RJ. 2016. The Biochemistry and Physiology of Mitochondrial Fatty Acid β-Oxidation and Its Genetic Disorders. Annu. Rev. Physiol. 78(1):23-44. https://doi.org/10.1146/annurev-physiol-021115-105045
2.
Ramsay RR, Gandour RD, van der Leij FR. 2001. Molecular enzymology of carnitine transfer and transport. Biochimica et Biophysica Acta (BBA) - Protein Structure and Molecular Enzymology. 1546(1):21-43. https://doi.org/10.1016/s0167-4838(01)00147-9
3.
McCann MR, George De la Rosa MV, Rosania GR, Stringer KA. 2021. L-Carnitine and Acylcarnitines: Mitochondrial Biomarkers for Precision Medicine. Metabolites. 11(1):51. https://doi.org/10.3390/metabo11010051
4.
Shekhawat PS, Matern D, Strauss AW. 2005. Fetal Fatty Acid Oxidation Disorders, Their Effect on Maternal Health and Neonatal Outcome: Impact of Expanded Newborn Screening on Their Diagnosis and Management. Pediatr Res. 57(5 Part 2):78R-86R. https://doi.org/10.1203/01.pdr.0000159631.63843.3e
5.
Adams SH, Hoppel CL, Lok KH, Zhao L, Wong SW, Minkler PE, Hwang DH, Newman JW, Garvey WT. 2009. Plasma Acylcarnitine Profiles Suggest Incomplete Long-Chain Fatty Acid β-Oxidation and Altered Tricarboxylic Acid Cycle Activity in Type 2 Diabetic African-American Women. The Journal of Nutrition. 139(6):1073-1081. https://doi.org/10.3945/jn.108.103754