Comparison with Particulate Material
The selectivity of Chromolith® Prep RP-18 endcapped monolithic columns is comparable to common RP-18 endcapped reversed phase columns. However, the high speed and high efficiency of these preparative columns make them excellent for separations of non-polar basic and acidic compounds as well as peptides.
In most cases, existing methods using particulate columns can be easily transferred to Chromolith® Prep. However, for some applications, it is worth optimizing the method in order to make use of the full potential of this enhanced technology.
Comparison of Chromolith® Prep RP-18 Endcapped 100-25 mm with Particulate Material: Separation of Oxime Derivates
The comparison of Chromolith® Prep RP-18 endcapped with a particulate Purospher RP-18 endcapped column (250-50 mm) shows that both separations have a similar resolution under the same chromatographic conditions. However, Chromolith® Prep RP-18 endcapped has better selectivity than Purospher RP-18 endcapped, as exhibited by the resolution of the additional isomer peak at 7 minutes.
Chromolith® Prep RP-18 endcapped 100-25 mm
| Column  | 
Chromolith® Prep RP-18 endcapped 100-25 mm | 
| Mobile Phase | 
A: Water + 0.05% TFA B: Acetonitrile + 0.05% TFA | 
| Gradient | 
 linear in 11 min up to 100% B 
 | 
| Flow rate | 
100 mL/min | 
| Detection  | 
UV 210 nm | 
| Sample | 
125 mg oxime derivates in 500 μL Acetonitrile | 
  | 
 
 
Separation of Hirudin (Filtrate of Crude Extract)
Without any sample preparation, the crude sample was injected directly onto the Chromolith® Prep RP-18 endcapped column. The separation took only 5 minutes, in which the target product was successfully isolated from the impurities.
Chromolith® Prep RP-18 endcapped 100-25 mm
| Column  | 
Chromolith® Prep RP-18 endcapped 100-25 mm | 
| Mobile Phase | 
A: Water + 0.1% Formic acid B: Acetonitrile 100% | 
| Gradient | 
Time [min] | 
%A | 
%B | 
| 0 | 
90 | 
10 | 
| 10 | 
70 | 
30 | 
| 10.1 | 
90 | 
10 | 
| Flow rate | 
60 mL/min | 
| Detection | 
UV 254 nm | 
| Sample | 
23 mg hirudin (filtrate of crude extract) injected in 5 mL solution | 
  | 
 
 
Separation of Dihydropyridines (Nifedipine, Nimodipine and Nisoldipine)
With the monolithic silica rod technology of Chromolith® Prep RP-18 endcapped columns, it is possible to speed up separations significantly while reducing back pressure.
Chromolith® Prep RP-18 endcapped 100-25 mm
| Column | 
Chromolith® Prep RP-18 endcapped 100-25 mm | 
| Mobile Phase | 
A: Water B: Acetonitrile | 
| Gradient | 
Time [min] | 
%A | 
%B | 
| 0 | 
80 | 
20 | 
| 8 | 
20 | 
80 | 
| 8.1 | 
80 | 
20 | 
| Flow rate | 
10 mL/min | 
| Detection | 
UV 224 nm | 
| Sample | 
90 mg mixture of nifedipine, nimodipine and nisoldipine | 
  | 
 
 
Chromolith® Prep Product Range
The ready-to-use columns are cladded with a polymeric material (PEEK), and can be connected directly to any HPLC system.
Product
 | 
Ordering Number
 | 
| Chromolith® Prep Si | 
125251 | 
| Chromolith® Prep RP-18-e | 
125252 | 
| Chromolith® Prep Guard Cartridge Si | 
125260 | 
| Chromolith® Prep Guard Cartridge RP-18e | 
125261 | 
| Chromolith® Prep Sealing Set (2 O-rings) | 
125254 | 
| Chromolith® Prep Tool Set (1 mounting tool filter, 1 mounting tool, 1 hook wrench) | 
125255 | 
| Chromolith® Prep Endcap Set (1 inlet cap complete, 1 outlet cap) | 
125256 | 
| Chromolith® Prep Frit Set (10 frits) | 
125257 | 
| Chromolith® Prep 25 mm Column Couple | 
125259 |