Skip to Content
Merck

MCGBL4S03

Millidisk® Barrier Cartridge Media: Durapore® 0.22 µm

Chemistry: Hydrophobic/Hydrophilic Polyvinylidene Fluoride (PVDF)

Synonym(s):

Millidisk® Barrier Filter 0.22 µm hydrophilic/phobic, Hydrophilic/Hydrophobic Durapore® 0.22 µm, Millidisk® Barrier 40 Cartridge

Sign In to View Organizational & Contract Pricing.

Select a Size

Change View

About This Item

eCl@ss:
32031690
UNSPSC Code:
23151806
Technical Service
Need help? Our team of experienced scientists is here for you.
Let Us Assist


Product Name

Millidisk® Barrier Cartridge Media: Durapore® 0.22 µm, Chemistry: Hydrophobic/Hydrophilic Polyvinylidene Fluoride (PVDF)

Quality Segment

material

PVDF , polysulfone , polysulfone support, silicone seal

description

hydrophobic/hydrophilic

sterility

non-sterile

sterilization compatibility

autoclavable compatible
steam-in-place compatible

product line

Millidisk® Barrier 40

feature

hydrophilic, hydrophobic

manufacturer/tradename

Millidisk®

parameter

0.35 bar max. differential pressure (5 psid) at 123 °C (Forward), 0.35 bar max. inlet pressure (5 psi) at 123 °C (Forward), 0.7 bar max. differential pressure (10 psid) at 25 °C (Reverse), 0.7 bar max. inlet pressure (10 psi) at 25 °C (Reverse), 1.7 bar max. differential pressure (25 psid) at 80 °C (Forward), 1.7 bar max. inlet pressure (25 psi) at 80 °C (Forward), 4.1 bar max. differential pressure (60 psid) at 25 °C (Forward), 4.1 bar max. inlet pressure (60 psi) at 25 °C (Forward), 7.6 L/min flow rate at 1 bar (ΔP)

technique(s)

gas filtration: suitable

filtration area

2000 cm2

impurities

≤0.5 EU/mL bacterial endotoxins (LAL test, Aqueous extraction), <0.5 EU/mL USP bacterial endotoxins (LAL test, sample aqueous extraction)

matrix

Durapore®

pore size

0.22 μm

input

sample type liquid

bubble point

≥1280 mbar (18.5 psig), nitrogen with 70/30% IPA/water at 23 °C

cartridge code

not applicable

General description

Stack Design: 40 Stak
Device Configuration: Cartridge

Packaging

Double Easy-Open bag

Preparation Note

Sterilization Method
4 SIP cycles of 60 min @ 135 °C; 4 autoclave cycles of 60 min @ 126 °C
This product was manufactured with a Durapore® membrane which meets the criteria for a "non-fiber releasing" filter as defined in 21 CFR 210.3 (b) (6).

Analysis Note

Bacterial Retention
Samples were quantitatively retentive of a minimum Brevundimonas diminuta challenge concentration of 1 x 10⁷ CFU/cm² using ASTM® F838 methodology.

Other Notes

Directions for Use

  • Organism Retention: Microorganism
  • Mode of Action: Filtration (size exclusion)
  • Application: BioProcessing
  • Intended Use: Reduction or removal of microorganism/bioburden
  • Instructions for Use: Please refer Wetting Instructions for Filter Units with Durapore® Membrane user guide
  • Storage Statement: Store in dry location
  • Disposal Statement: Dispose of in accordance with applicable federal, state and local regulations.

These products are manufactured in a facility which adheres to FDA Good Manufacturing Practices.

Legal Information

ASTM is a registered trademark of American Society for Testing and Materials
Durapore is a registered trademark of Merck KGaA, Darmstadt, Germany
MILLIDISK is a registered trademark of Merck KGaA, Darmstadt, Germany




Certificates of Analysis (COA)

Search for Certificates of Analysis (COA) by entering the products Lot/Batch Number. Lot and Batch Numbers can be found on a product’s label following the words ‘Lot’ or ‘Batch’.

Already Own This Product?

Find documentation for the products that you have recently purchased in the Document Library.

Visit the Document Library