製品名
Fluoride Test, photometric, 0.02-2.00 mg/L (F), Spectroquant®
product line
Spectroquant®
Quality Segment
usage
sufficient for 250 tests
specific analyte(s)
fluoride
measuring range
0.02-2.00 mg/L (F)
technique(s)
photometry: suitable
application(s)
cannabis testing
disposal drainage water
drinking water
electroplating
environmental
ground and surface water
mineral water
seawater
wastewater
compatibility
for use with Spectroquant® Move 100, for use with Spectroquant® Nova 60 A, for use with Spectroquant® Prove 100 (plus), for use with Spectroquant® Prove 300 (plus), for use with Spectroquant® Prove 600 (plus)
detection method
photometric (SPADNS method)
storage temp.
15-25°C
General description
Method: photometric, SPADNS 0.02 - 2.00 mg/l F Spectroquant®
This Spectroquant® Fluoride test allows the accurate quantification of fluoride ions in various waters. Fluoride ions occur on earth in several minerals, but are only present in trace quantities in bodies of water in nature. Fluoride ions react with part of a red zirconium-dye solution to form a colorless complex, which in turn results in a bleaching of the red color (SPADNS method). The concentration of the unconsumed zirconium-dye solution is subsequently determined photometrically.
The method is analogous to APHA 4500-F- D.
All our Cell and Reagent Test Kits are equipped with the unique Live ID (2D barcode) which allows seamless method recognition and contains essential information such as lot number, expiry date, and automatic calibration updates.
The Spectroquant® Reagent Test Kits contain highly stable, ready-to-use reagent mixtures to perform the analysis according to the procedure described in the accompanying instruction leaflet.
This Spectroquant® Fluoride test allows the accurate quantification of fluoride ions in various waters. Fluoride ions occur on earth in several minerals, but are only present in trace quantities in bodies of water in nature. Fluoride ions react with part of a red zirconium-dye solution to form a colorless complex, which in turn results in a bleaching of the red color (SPADNS method). The concentration of the unconsumed zirconium-dye solution is subsequently determined photometrically.
The method is analogous to APHA 4500-F- D.
All our Cell and Reagent Test Kits are equipped with the unique Live ID (2D barcode) which allows seamless method recognition and contains essential information such as lot number, expiry date, and automatic calibration updates.
The Spectroquant® Reagent Test Kits contain highly stable, ready-to-use reagent mixtures to perform the analysis according to the procedure described in the accompanying instruction leaflet.
Legal Information
SPECTROQUANT is a registered trademark of Merck KGaA, Darmstadt, Germany
Still not finding the right product?
Explore all of our products under Fluoride Test
注意喚起語
Warning
Hazard Classifications
Eye Irrit. 2 - Met. Corr. 1 - Skin Irrit. 2 - STOT SE 3
標的臓器
Respiratory system
引火点 (F)
Not applicable
引火点(°C)
Not applicable
ハザードコード
保管分類
12 - Non Combustible Liquids
一体何が起きているの?
WGK 3