grade
analytical standard
Quality Segment
agency
according to Ph. Eur. stray light testing
product line
Certipur®
technique(s)
UV/Vis spectroscopy: suitable
application(s)
environmental
industrial qc
format
mixture
storage temp.
15-25°C
General description
for stray light testing according to Ph Eur Certipur®
Application
UV-VIS Standard 3: Sodium iodide solution as a reference material for checking the stray light spectrophotometers in accordance with European Pharmacopoeia (Ph. Eur.) chapter 2.2.25 and American Society for Testing and Materials (ASTM) E 387 and E 925.
Features and Benefits
- It is a 10 g/L solution of sodium iodide in H2O
- Conveniently packed in 10 mL glass ampules. Each box includes 3 single-use ampules
- Stable solutions in glass ampoules
- Accreditation according to ISO 17025
- Traceability to standard reference material from the National Institute of Standards and Technology (NIST) and Physical-Technical Institute (PTB)
- Includes a certificate of analysis (CoA)
- Reliable and reproducible results
Legal Information
CERTIPUR is a registered trademark of Merck KGaA, Darmstadt, Germany
signalword
Danger
hcodes
Hazard Classifications
Repr. 1B
wgk
WGK 2
保管分類
6.1D - Non-combustible acute toxic Cat.3 / toxic hazardous materials or hazardous materials causing chronic effects
適用法令
試験研究用途を考慮した関連法令を主に挙げております。化学物質以外については、一部の情報のみ提供しています。 製品を安全かつ合法的に使用することは、使用者の義務です。最新情報により修正される場合があります。WEBの反映には時間を要することがあるため、適宜SDSをご参照ください。
Please refer to KIT Component information
pdsc
Please refer to KIT Component information
prtr
Please refer to KIT Component information
fsl
Please refer to KIT Component information
ishl_indicated
Please refer to KIT Component information
ishl_notified
Please refer to KIT Component information
cart
キットコンポーネントの情報を参照してください
jan