で組織・契約価格をご覧ください。
サイズを選択してください
表示を変更する
この商品について
製品名
水酸化ナトリウム 溶液, c(NaOH) = 0.2 mol/l (0.2 N), Titripur®
vapor pressure
3 mmHg ( 37 °C)
Quality Segment
product line
Titripur®
form
liquid
quality
Analyzed in our ISO 17025 accredited QC lab
reaction suitability
reaction type: Acid-base reactions
packaging
Titripac® of
concentration
0.2 M
technique(s)
titration: suitable
pH
13 (20 °C in H2O)
density
1.01 g/cm3 at 20 °C
storage temp.
15-25°C
SMILES string
[OH-].[Na+]
InChI
1S/Na.H2O/h;1H2/q+1;/p-1
InChI key
HEMHJVSKTPXQMS-UHFFFAOYSA-M
Features and Benefits
Digital Titration
ChemiSphere™ App for Digital Data Transfer to OMNIS Titration Instruments from Metrohm
Discover the ChemiSphere™ App for secure, rapid data transfer to OMNIS titrators, enhancing efficiency in the titration workflow with Titripur® and Aquastar® reagents. ChemiSphere App for Digital Reagent Data Transfer
SmartChemicals for Digital Data Transfer to Mettler Toledo Titration Instruments
Explore digital data transfer with our SmartChemicals for Mettler Toledo titrators—fast, easy, and safe, accelerating the titration process with Titripur®, Aquastar® reagents, and Certipur® and Aquastar® standards. Supelco® SmartChemicals for Titration
ChemiSphere™ App for Digital Data Transfer to OMNIS Titration Instruments from Metrohm
Discover the ChemiSphere™ App for secure, rapid data transfer to OMNIS titrators, enhancing efficiency in the titration workflow with Titripur® and Aquastar® reagents. ChemiSphere App for Digital Reagent Data Transfer
SmartChemicals for Digital Data Transfer to Mettler Toledo Titration Instruments
Explore digital data transfer with our SmartChemicals for Mettler Toledo titrators—fast, easy, and safe, accelerating the titration process with Titripur®, Aquastar® reagents, and Certipur® and Aquastar® standards. Supelco® SmartChemicals for Titration
Accreditation:
This volumetric solution is analyzed by our calibration laboratory D-K-15185-01-00 which is accredited according to DIN EN ISO/IEC 17025 for analysis of amount-of-substance concentrations in volumetric solutions by DAkkS (Deutsche Akkreditierungsstelle - German National Accreditation Body). The accreditation certificate can be found at www.sigmaaldrich.com/ISO17025.
This volumetric solution is analyzed by our calibration laboratory D-K-15185-01-00 which is accredited according to DIN EN ISO/IEC 17025 for analysis of amount-of-substance concentrations in volumetric solutions by DAkkS (Deutsche Akkreditierungsstelle - German National Accreditation Body). The accreditation certificate can be found at www.sigmaaldrich.com/ISO17025.
Packaging
SKU-pack size numbers ending with a "3" are SmartChemicals with an RFID tag on the label for seamless data transfer to the instrument, e.g. xxxxxx1003 or xxxxxxx 4003. More information on www.sigmaaldrich.com/SmartChemicals
Titripac® packaging - all SKU-pack size numbers ending "4000, 4003, 9010, 9013" - more information on www.sigmaaldrich.com/Titripac
Titripac® packaging - all SKU-pack size numbers ending "4000, 4003, 9010, 9013" - more information on www.sigmaaldrich.com/Titripac
Analysis Note
Form liquid
Amount-of-substance concentration 0.1990 - 0.2010 mol/L
Measurement uncertainty ± 0.0006 mol/L
Traceability NIST SRM
The concentration is determined by volumetric titration and refers to 20°C.
The amount-of-substance concentration of this volumetric solution is traceable to a primary standard reference material (SRM) from the National Institute of Standards and Technology, Gaithersburg, USA (NIST SRM 84 potassium hydrogen phthalate) by means of volumetric standard potassium hydrogen phthalate (article number 1.02400), certified reference material according to ISO 17034, analyzed by our accredited calibration laboratory of Merck KGaA, Darmstadt, Germany according to DIN EN ISO/IEC 17025. The uncertainty is expressed as expanded measurement uncertainty with a coverage factor k=2 covering a confidence level of 95%.
Note: The titer is a correction factor to correct for variations of the volumetric solution, the titration equipment, the temperature and other laboratory conditions. For correct titration results it is recommended to determine a titer with the laboratory specific equipment and under laboratory specific conditions directly after opening a new bottle and at regular time intervals.
Amount-of-substance concentration 0.1990 - 0.2010 mol/L
Measurement uncertainty ± 0.0006 mol/L
Traceability NIST SRM
The concentration is determined by volumetric titration and refers to 20°C.
The amount-of-substance concentration of this volumetric solution is traceable to a primary standard reference material (SRM) from the National Institute of Standards and Technology, Gaithersburg, USA (NIST SRM 84 potassium hydrogen phthalate) by means of volumetric standard potassium hydrogen phthalate (article number 1.02400), certified reference material according to ISO 17034, analyzed by our accredited calibration laboratory of Merck KGaA, Darmstadt, Germany according to DIN EN ISO/IEC 17025. The uncertainty is expressed as expanded measurement uncertainty with a coverage factor k=2 covering a confidence level of 95%.
Note: The titer is a correction factor to correct for variations of the volumetric solution, the titration equipment, the temperature and other laboratory conditions. For correct titration results it is recommended to determine a titer with the laboratory specific equipment and under laboratory specific conditions directly after opening a new bottle and at regular time intervals.
Legal Information
AQUASTAR is a registered trademark of Merck KGaA, Darmstadt, Germany
CERTIPUR is a registered trademark of Merck KGaA, Darmstadt, Germany
ChemiSphere is a trademark of Merch KGaA, Darmstadt, Germany
Supelco is a registered trademark of Merck KGaA, Darmstadt, Germany
Titripac is a registered trademark of Merck KGaA, Darmstadt, Germany
Titripur is a registered trademark of Merck KGaA, Darmstadt, Germany
Still not finding the right product?
Explore all of our products under 水酸化ナトリウム 溶液
signalword
Warning
Hazard Classifications
Eye Irrit. 2 - Met. Corr. 1 - Skin Irrit. 2
hcodes
保管分類
12 - Non Combustible Liquids
試験成績書(COA)
製品のロット番号・バッチ番号を入力して、試験成績書(COA) を検索できます。ロット番号・バッチ番号は、製品ラベルに「Lot」または「Batch」に続いて記載されています。