SeQuant® ZIC®-HILIC HPLC Columns |
Request Information |
SeQuant® ZIC®-HILIC HPLC columns have documented excellent selectivity for separations of polar hydrophilic compounds in HPLC and LC-MS methods. These columns carry densely bonded, truly zwitterionic functional groups with a charge balance of 1:1. Separation is achieved by hydrophilic partitioning combined with weak ionic interactions resulting in maximum selectivity, high loadability and easy optimization of methods. The comprehensive SeQuant® ZIC®-HILIC product range includes classical analytical and capillary columns, as well as semi-preparative formats. All columns are available with 3 or 5 µm particles, and pore sizes of 100 or 200 Å.
Its favorable retention and robust performance have made SeQuant® ZIC®-HILIC one of the most popular HILIC columns throughout the world. The columns have been used in numerous methods for a variety of compounds, such as acids and bases, anions and cations, carbohydrates, metabolites, metal complexes, amino acids, peptides, protein digests and many more. Alternative choices for separation of polar hydrophilic compounds include SeQuant® ZIC®-cHILIC with complementary selectivity, and SeQuant® ZIC®-pHILIC for higher pH stability.
A group of distinguished Japanese scientists performed comprehensive screening to compare the performance of 15 different HILIC columns. Their conclusion was that the selectivity of SeQuant® ZIC®-HILIC was higher and more versatile than that of all other stationary phases.
“In particular, the zwitterionic phase ZIC®-HILIC showed better selectivity for the test samples than the other phases.”
Kawachi et al., Journal of Chromatography A, Vol. 1218, Issue 35, 2 September 2011, Pages 5903–5919
HILIC or hydrophilic interaction liquid chromatography is a high-performance liquid chromatographic (HPLC) technique for separation of polar and hydrophilic compounds. Simply put, HILIC is a normal-phase type of separation but uses reversed-phase type eluents. Thus, it unites the benefits of aqueous mobile phases with separation based on normal phase principles in which more polar compounds respond with higher retention. This results in more efficient separation. Furthermore, HILIC is easy to combine with a wide range of detection methods, and boasts improved sensitivity with some of these techniques, such as electrospray ionization mass spectrometry (ESI-MS).
Among numerous other applications, the use of SeQuant® ZIC®-HILIC columns has been recommended by the US FDA (United States Food and Drug Administration) in its methods for the analysis of melamine ad related potential contaminants. Learn more about melamine analysis in the pharmaceutical industry.
Gradient | Time | A: 95% acetonitrile in 0.1% aqueous formic acid | B: 50% acetonitrile in 20 mM aqueous ammonium formate | Flow rate [mL/min] |
---|---|---|---|---|
0 | 100 | 0 | 0.4 | |
4.2 | 100 | 0 | 0.4 | |
8.0 | 65 | 35 | 0.4 | |
8.5 | 65 | 35 | 0.4 | |
9.0 | 25 | 75 | 0.4 | |
11.0 | 25 | 75 | 0.4 | |
11.2 | 100 | 0 | 0.6 | |
13.0 | 100 | 0 | 0.6 | |
14.0 | 100 | 0 | 0.4 | |
ZIC®-HILIC 150 x 2.1 mm, 5 μm, 200 Å | ||||
Column | ZIC®-HILIC 150 x 2.1 mm, 5 μm, 200 Å | |||
Detection | MS/MS, -ESI and +ESI | |||
Temperature | 30°C | |||
Inj. volume | 5 μL | |||
Sample | Standard in mobile phase (7 ng/mL) equivalent to 1 μg/g |
|||
Courtesy of | David N. Heller, FDA Center for Veterinary Medicine, Laurel, MD 20708 USA | |||
Reference | David N. Heller and Cristina B. Nochetto, Rapid Commun. Mass Spectrom., 22 (2008) 3624-3632. DOI: 10.1002/rcm.3779 |
|||
Column | ZIC®-HILIC 150 x 4.6 mm, 3.5 μm, 100 Å | |
---|---|---|
Mobile Phase | 80% | acetonitrile |
20% | ammonium acetate in water, 50 mM pH adjusted to 4.5 with formic acid | |
Flow rate | 0.75 mL/min | |
Detection | Refractive index, cell 9 μL, 40°C | |
Temperature | 40°C | |
Inj. volume | 50 μL | |
Gradient | 1. 2-Amino-5-Hydroxy benzoic acid 2. Tryptophan 3. Isoleucine 4. Methionine 5. Lycine 6. Threonine |
100 ppm 100 ppm 100 ppm 100 ppm 100 ppm 100 ppm |
Courtesy of | Gora Sharangi, Merck India Application Lab | |
Column | ZIC®-HILIC 100 x 2.1 mm, 3.5 μm,100 Å (1.50444.0001) |
|
---|---|---|
Mobile Phase | A: | 80% (v/v) acetonitrile |
B: | 20% (v/v) 100 mM NH4Ac buffer pH 4.5 | |
Flow rate | MS-ESI- with SIM at 117.2 m/z Drying-gas flow rate Drying-gas temperature Capillary voltage |
10 L/min 300°C 3.0 kV |
Detection | Refractive index, cell 9 μL, 40°C | |
Inj. volume | 4 μL | |
Sample | Precipitation-treated plasma sample containing 0.137 mol/L MMA | |
Reference | H-A Lakso, P. Appelblad, J. Schneede Clin. Chem., 54 (2008) 2028 DOI: 10.1373/clinchem.2007.101253 |
|
SeQuant® Column | Functional group | Features | Base particle | Particle size | Pore size | Column types |
---|---|---|---|---|---|---|
SeQuant® ZIC®-HILIC |
Sulfobetaine |
Well-known for high-performance selectivity and robustness | silica | 3.5 µm, 5 µm |
100 Å, 200 Å | analytical, capillary, semi-prep, guards |
Merck offers an extensive range of services, information and tools to support you in your analysis. Use the links below to learn more.