Optiseal Sanitary Cartridge Housings

Recommended Products
Overview
Specifications
Ordering Information
Documentation
Related Products & Applications
Product Families
|   
 Durapore® 0.1 µm and 0.22 µm Optiseal® Cartridge FiltersOptiseal cartridge filters feature a revolutionary cartridge-to-housing sealing mechanism and pleated configuration, resulting in maximum sterility assurance and economical performance.Learn More >> | 
|   
 Hydrophobic Durapore® 0.22 µm Optiseal® Cartridge FiltersReliable filters for the sterile filtration of gases and liquids. Featuring a unique cartridge-to-housing sealing mechanism, these Optiseal Durapore cartridges offer maximum sterility assurance.Learn More >> | 
|   
 Durapore® 0.45 µm Optiseal® Cartridge FiltersOptiseal cartridge filters feature a revolutionary cartridge-to-housing sealing mechanism and pleated configuration, resulting in maximum sterility assurance and economical performance.Learn More >> | 
|   
 Optiseal Cartridge Filters with Hydrophobic Aervent MembraneOptiseal filters with Aervent membrane are designed for sterilizing applications, removing particles and microorganisms from gases and solvents.Learn More >> | 
|   
 Charged Durapore® 0.22 µm Optiseal® Cartridge FiltersOptiseal® hydrophilic cartridge filters with charged Durapore 0.22 µm membrane are sterilizing-grade filters manufactured from PVDF membrane modified to give it a net positive charge.Learn More >> | 
Related Products By: Application Facete
| Dialysis and Filtration | 
Related Products By: Brand Facete
| Optiseal® | 
Categories
| Biopharmaceutical Manufacturing > Downstream Processing > Multi-use Processing Systems > Filter Housings | 
Description
The Optiseal housing is designed for sterile filtration of aqueous solutions and gases using Optiseal cartridge filters. All wetted metal surfaces are constructed of 316L stainless steel and are electropolished, providing excellent durability and maximum corrosion resistance. 
The housing features easy-to-clean vents, drains, and connections, allowing complete sterilization. Vent-insulating rings are included for added safety. 
The housing accommodates a double O-ring cartridge configuration, allowing a positive cartridge seal and preventing bacterial and particulate bypass . The Optiseal housing is designed to engage hold-down tabs on the Optiseal cartridge, which lock the cartridge in place when the housing is fully assembled.
Cartridges Used
All Optiseal cartridge filters
Dimensions
| Inlet to Outlet, cm (in) | Length, cm (in) | Weight, kg (lb) | |
| In-line housing | 20.3 cm (8 in.) | 20.3 cm (8 in.) | 1.25 kg (2.75 lb.) | 
| T-line housing | 17.7 cm (7 in.) | 19.6 cm (7.7 in.) | 1.25 kg (2.75 lb.) | 
| Materials of Construction | ||
| Housing | 316L stainless steel | |
| Clamp | 304 stainless steel | |
| Seal | Silicone (SI) | |
| Vent Insulators | EPDM | |
| Maximum Operating Pressure, bar (psig) | 10 bar (145 psi) | |
| Surface Finish | ||
| Wetted Surfaces | Fine electropolished sanitary finish, Ra <0.5 µm (20 µin) | |
| Exterior Surfaces | Sanitary finish Ra <0.8 µm (32 µin) | |
| Connections | ||
| Inlet/Outlet | 1-1/2 in. Sanitary Flange | |
| Cartridge | Double O-rings (2-123) with hold-down tabs | |
| Dome to Base - In-line Housing | 3 in. O-ring (2-335) | |
| Dome to Base - T-line Housing | 3 in. O-ring (2-335) | |
| Mounting | Can be supported by braced rigid tubing connections | |
| Sterilization | By providing improved heat transfer, the steam penetration holes in the housing base result in a qualified steam sterilization cycle of 30 minutes at 135 °C | |

 


 
  

 
 